Endress+Hauser
Endress+Hauser Water hardness analyzer Liquiline System CA80HA
REQUEST A QUOTE
Measurement range | 0 to 80 mg/l CaCO3 0 to 80 mg/l with dilution function to maximum 16 to 1600 mg/l CaCO3 |
---|---|
Process temperature | 4 to 40 °C (39 to 104 °F) |
Process pressure | Unpressurized |
Measuring principle | Colorimetric |
Characteristic | Process analyzer for hardness in aqueous solutions Comply with standard colorimetric measuring principle - phthalein purple method |
Size | Housing (open version): 793 x 530 x 417 mm 31.22 x 20.87 x 16.42 in Housing (closed version): 793 x 530 x 463 mm 31.22 x 20.87 x 18.23 in Housing with base: 1723 x 530 x 463 mm 67.83 x 20.87 x 18.23 in |
Design | Open design, cabinet and stand housing High-Performance plastic ASA-PC, additional stand coated steel |
Process temperature | 4 to 40 °C (39 to 104 °F) |
Ambient temperature | 5 to 40 °C (41 to 104 °F) |
Process pressure | Unpressurized |
Sample flow rate | Min. 5 ml/min (0.17 fl.oz/min) |
Consistency of the sample | Low solid content (< 50NTU), aqueous homogeneous |
Application | Total hardness monitoring in reverse osmosis systems inlets, after ion exchange systems Monitoring of water preparation in processes and drinking water |
Power supply | 100 ... 120 VAC / 200 ... 240 VAC ± 10% 24 VDC ± 10% 50 ± 1 oder 60 ± 1,2 Hz |
Output / communication | 2x 0/4 ... 20 mA Webserver, Modbus, Ethernet/IP, Profibus DP |
Input | 1 or 2 measuring channel 1 ... 4 digital sensor inputs for sensors with Memosens protocol (optional) |
Measurement range | 0 to 80 mg/l CaCO3 0 to 80 mg/l with dilution function to maximum 16 to 1600 mg/l CaCO3 |
Consumables | Reagents and standard solutions CY80HA as well as cleaner CY800 are necessary for the operation Regular maintenance is done with the parts of the maintenance kit CAV800 |
Returns Policy