Endress+Hauser TN analyzer Liquiline System CA80TN

Write a Review

REQUEST A QUOTE

Endress+Hauser TN analyzer Liquiline System CA80TN
Measurement range0 to 10 mg/l N
0 to 50 mg/l N
0 to 200 mg/l N
Process temperature4 to 40 °C (39.2 to 104 °F)
Process pressureAt atmospheric pressure, < 0.2 bar
Measuring principleUV photometric
CharacteristicProcess analyzer for total nitrogen in aqueous solutions
UV photometric method following HJ636-2012
SizeHousing (open version):
793 x 530 x 417 mm
31.22 x 20.87 x 16.42 in
Housing (closed version):
793 x 530 x 463 mm
31.22 x 20.87 x 18.23 in
Housing with base:
1723 x 530 x 463 mm
67.83 x 20.87 x 18.23 in
DesignOpen design, cabinet and stand housing
High-Performance plastic ASA-PC, additional stand coated steel
Process temperature4 to 40 °C (39.2 to 104 °F)
Ambient temperature5 to 40 °C (41 to 104 °F)
Process pressureAt atmospheric pressure, < 0.2 bar
Sample flow rateMin. 15 ml/min (0.51 fl.oz/min)
Consistency of the sampleAqueous, homogeneous
ApplicationMonitoring of wastewater treatment plant outlet
Monitoring of the surface water quality
Monitoring of industrial water treatment
Monitoring and optimization of industrial wastewater treatment plants
Power supply100 to 120 VAC / 200 to 240 VAC ± 10%
50 ± 1 or 60 ± 1,2 Hz
Output / communication2x 0/4 to 20 mA
Modbus RS485/TCP (optional)
Webserver (optional)
Ethernet/IP
Profibus DP
Alarm relay
Input1 measuring channel
1 to 4 digital sensor inputs for sensors with Memosens protocol (optional)
Measurement range0 to 10 mg/l N
0 to 50 mg/l N
0 to 200 mg/l N
ConsumablesReagents and standard solutions CY80TN as well as cleaner CY800 are necessary for the operation
Regular maintenance is done with the parts of the maintenance kit CAV880
Returns Policy