Accuracy | Standard acc. to ASTM E-230 Special acc. to ASTM E-230 |
---|---|
Response time | depending on configuration 63% rt = 15 s |
Max. process pressure (static) | at 20°C: 100 bar (1.450psi) depends on configuration |
Operating temperature range | Type T: -270 °C ...370 °C (-454 °F ...698 °F) Type J: -210 °C ...720 °C (-346 °F ...1.328 °F) Type E: -270 °C ...820 °C (-454 °F ...1.508 °F) Type K: -270 °C ... 1.150 °C (-454 °F ...2.102 °F) Type N: -270 °C ... 1.150 °C (-454 °F ...2.102 °F) |
Max. immersion length on request | up to 24'' (609 mm) others on request |
Measuring principle | Thermocouple |
Characteristic / Application | Industrial US style modular temperature assembly threaded, weld-in, and socket weld process connection with extension incl. thermowell |
Thermowell / protection tube | bar stock (drilled) |
Insert / probe | mineral insulated (MI), flexible |
Outer diameter protection tube / Insert | 3/4'' (19,05 mm) 7/8'' (22,23 mm) 1'' (25,40 mm) 17/16'' (26,99 mm) 0,63'' ...1,31'' (15,88 mm ... 33,40 mm) |
Max. immersion length on request | up to 24'' (609 mm) others on request |
Material protection tube/ thermowell | 316/316L others on request |
Optional coating | Available on request |
Process connection | male thread: NPT1/2" NPT3/4" NPT1" weld in version socket weld version |
Tip shape | straight stepped tapered |
Surface roughness Ra | 32 μin (0.80 μm) |
Operating temperature range | Type T: -270 °C ...370 °C (-454 °F ...698 °F) Type J: -210 °C ...720 °C (-346 °F ...1.328 °F) Type E: -270 °C ...820 °C (-454 °F ...1.508 °F) Type K: -270 °C ... 1.150 °C (-454 °F ...2.102 °F) Type N: -270 °C ... 1.150 °C (-454 °F ...2.102 °F) |
Max. process pressure (static) | at 20°C: 100 bar (1.450psi) depends on configuration |
Accuracy | Standard acc. to ASTM E-230 Special acc. to ASTM E-230 |
Response time | depending on configuration 63% rt = 15 s |
Integration head transmitter | yes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS) |
Ex - approvals | transmitter only FM IS CSA IS FM/CSA IS |
Certification | SIL (transmitter only) |