Endress+Hauser SS2100I SS2100i gas analyzer

Write a Review

REQUEST A QUOTE

Endress+Hauser SS2100I SS2100i gas analyzer
Measured VariablesConcentration
Cell Pressure
Cell Temperature
Hazardous area approvalsIECEx / ATEX / CNEx / KC / CCOE / CML Zone 1
Measuring principleTDLAS
Product HeadlineAn advanced gas analyzer for hydrogen sulfide (H2S) measurement. The SS2100i uses tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
Channels1
Analyte and Measurement rangesH2S (Hydrogen Sulfide):
0 to 10 ppmv
0 to 1000 ppmv
0 to 5000 ppmv
0 to 5% volume

Measured VariablesConcentration
Cell Pressure
Cell Temperature
Ambient Temperature range-20 to 50°C (-4 to 122°F)
Optional (i-2 only): -10 to 60°C (14 to 140°F)
Operating Pressure rangeInlet Pressure: 140 to 350 kPa (20 to 50 psig)
Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)
Analyzer Wetted materials316L stainless steel
FKM O-Rings
Glass
Power supply120 or 240 VAC ± 10 %, 50 to 60 Hz; 300W
CommunicationAnalog Output: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load (measurement value)
Analog Input: One 4-20mA Isolated, 1200 ohms at 24 VDC max load (process pressure)
Serial: RS232C and Ethernet
Protocol: Modbus Gould RTU or Daniel RTU or ASCII
Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
Digital Inputs: 2, Flow Alarm, Validation Request
Housing materialsElectronics: Copper-free Aluminum
Laser/Cell Enclosure: Copper-free Aluminum (i-2 only)
Sample System Enclosure: 304 or 316 stainless steel
Hazardous area approvalsIECEx / ATEX / CNEx / KC / CCOE / CML Zone 1
Degree of protectionIP66
Product safetyCE
Metrological approvalsPAC China, Kazakhstan (i-1 only)
Measuring principleTDLAS
Product HeadlineAn advanced gas analyzer for moisture (H2O) measurement. The SS2100i uses tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
Channels1
Analyte and Measurement rangesH2O (Moisture):
0 to 10 ppmv
0 to 100 ppmv
0 to 50 ppmv
0 to 5000 ppmv
Measured VariablesConcentration
Moisture Dew Point
Cell Pressure
Cell Temperature
Ambient Temperature range-20 to 50°C (-4 to 122°F)
Optional (i-2 only): -10 to 60°C (14 to 140°F)
Operating Pressure rangeInlet Pressure: 140 to 350 kPa (20 to 50 psig)
Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)
Analyzer Wetted materials316L stainless steel
FKM O-Rings
Glass
Power supply120 or 240 VAC ± 10 %, 50 to 60 Hz; 300W
CommunicationAnalog Output: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load (measurement value)
Analog Input: One 4-20mA Isolated, 1200 ohms at 24 VDC max load (process pressure)
Serial: RS232C and Ethernet
Protocol: Modbus Gould RTU or Daniel RTU or ASCII
Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
Digital Inputs: 2, Flow Alarm, Validation Request
Housing materialsElectronics: Copper-free Aluminum
Laser/Cell Enclosure: Copper-free Aluminum (i-2 only)
Sample System Enclosure: 304 or 316 stainless steel
Hazardous area approvalsIECEx / ATEX / CNEx / KC / CCOE / CML Zone 1
Degree of protectionIP66
Product safetyCE
Metrological approvalsPAC China, Kazakhstan (i-1 only)
Measuring principleTDLAS
Product HeadlineAn advanced gas analyzer for carbon dioxide (CO2) measurement. The SS2100i uses tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
Channels1
Analyte and Measurement rangesCO2 (Carbon Dioxide):
0 to 10 ppmv
0 to 1000 ppmv
0 to 5000 ppmv
0 to 5% volume
0 to 10% volume
0 to 20% volume
Measured VariablesConcentration
Cell Pressure
Cell Temperature
Ambient Temperature range-20 to 50°C (-4 to 122°F)
Optional (i-2 only): -10 to 60°C (14 to 140°F)
Operating Pressure rangeInlet Pressure: 140 to 350 kPa (20 to 50 psig)
Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)
Analyzer Wetted materials316L stainless steel
FKM O-Rings
Glass
Power supply120 or 240 VAC ± 10 %, 50 to 60 Hz; 300W
CommunicationAnalog Output: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load (measurement value)
Analog Input: One 4-20mA Isolated, 1200 ohms at 24 VDC max load (process pressure)
Serial: RS232C and Ethernet
Protocol: Modbus Gould RTU or Daniel RTU or ASCII
Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
Digital Inputs: 2, Flow Alarm, Validation Request
Housing materialsElectronics: Copper-free Aluminum
Laser/Cell Enclosure: Copper-free Aluminum (i-2 only)
Sample System Enclosure: 304 or 316 stainless steel
Hazardous area approvalsIECEx / ATEX / CNEx / KC / CCOE / CML Zone 1
Degree of protectionIP66
Product safetyCE
Metrological approvalsPAC China, Kazakhstan (i-1 only)
Measuring principleTDLAS
Product HeadlineAn advanced gas analyzer for ammonia (NH3) measurement. The SS2100i uses tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
Channels1
Analyte and Measurement rangesNH3 (Ammonia): 0 to 5 ppmv
Measured VariablesConcentration
Cell Pressure
Cell Temperature
Ambient Temperature range-20 to 50°C (-4 to 122°F)
Optional (i-2 only): -10 to 60°C (14 to 140°F)
Operating Pressure rangeInlet Pressure: 140 to 350 kPa (20 to 50 psig)
Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)
Analyzer Wetted materials316L stainless steel
FKM O-Rings
Glass
Power supply120 or 240 VAC ± 10 %, 50 to 60 Hz; 300W
CommunicationAnalog Output: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load (measurement value)
Analog Input: One 4-20mA Isolated, 1200 ohms at 24 VDC max load (process pressure)
Serial: RS232C and Ethernet
Protocol: Modbus Gould RTU or Daniel RTU or ASCII
Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
Digital Inputs: 2, Flow Alarm, Validation Request
Housing materialsElectronics: Copper-free Aluminum
Laser/Cell Enclosure: Copper-free Aluminum (i-2 only)
Sample System Enclosure: 304 or 316 stainless steel
Hazardous area approvalsIECEx / ATEX / CNEx / KC / CCOE / CML Zone 1
Degree of protectionIP66
Product safetyCE
Metrological approvalsPAC China, Kazakhstan (i-1 only)
Measuring principleTDLAS
Product HeadlineAn advanced gas analyzer for acetylene (C2H2) measurement. The SS2100i uses tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
Channels1
Analyte and Measurement rangesC2H2 (Acetylene):
0 to 5 ppmv (minimum)
0 to 3000 ppmv (maximum)
Measured VariablesConcentration
Cell Pressure
Cell Temperature
Ambient Temperature range-20 to 50°C (-4 to 122°F)
Optional (i-2 only): -10 to 60°C (14 to 140°F)
Operating Pressure rangeInlet Pressure: 140 to 350 kPa (20 to 50 psig)
Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)
Analyzer Wetted materials316L stainless steel
FKM O-Rings
Glass
Power supply120 or 240 VAC ± 10 %, 50 to 60 Hz; 300W
CommunicationAnalog Output: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load (measurement value)
Analog Input: One 4-20mA Isolated, 1200 ohms at 24 VDC max load (process pressure)
Serial: RS232C and Ethernet
Protocol: Modbus Gould RTU or Daniel RTU or ASCII
Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
Digital Inputs: 2, Flow Alarm, Validation Request
Housing materialsElectronics: Copper-free Aluminum
Laser/Cell Enclosure: Copper-free Aluminum (i-2 only)
Sample System Enclosure: 304 or 316 stainless steel
Hazardous area approvalsIECEx / ATEX / CNEx / KC / CCOE / CML Zone 1
Degree of protectionIP66
Product safetyCE
Metrological approvalsPAC China, Kazakhstan (i-1 only)
Returns Policy

Related Products

Top