Endress+Hauser Proline Prosonic Flow 92F ultrasonic flowmeter

Write a Review

REQUEST A QUOTE

Endress+Hauser Proline Prosonic Flow 92F ultrasonic flowmeter
Max. measurement errorVolume flow (standard):
- ±0.5 % o.r. for 0.5 to 10 m/s (1.6 to 33 ft/s)
Volume flow (option):
- ±0.3 % o.r. for 0.5 to 10 m/s (1.6 to 33 ft/s)
Measuring range0.5 to 10 m/s (1.6 to 33 ft/s)
Medium temperature range-40 to +150 °C (-40 to +302 °F)
-40 to +200 °C (-40 to +392 °F) optional
Max. process pressurePN 40 / ASME Cl. 300 / JIS 20K
Wetted materialsSensor:
A351-CF3M (DN25 to 100)
1.4404/TP316/TP316L or A106 GrB (DN150 to 300)
Transducer: 1.4404/316/316L
Flanges : 1.4404/316/316L or A105/1.0432
Measuring principleUltrasonic flow
Product headlineHigly accurate, loop-powered flowmeter with innovative parallel path design for minimum inlet runs.
Inline device for homogeneous conductive and non-conductive liquids in the chemical and petrochemical industry.
Sensor featuresSafe design for process industries – international hazardous area approvals. No additional pressure loss – full-bore design. Process transparency – diagnostic capability.
Full compliance according to NACE MR0175 and MR010. Nominal diameter: DN 25 to 300 (1 to 12"). Medium temperature: –40 to 200 °C (–40 to 392 °F).
Transmitter featuresEasy installation and reduced installation costs – loop-powered transmitter. Fully industry compliant – IEC/ATEX/FM/CSA/JPN/NEPSI. Automatic recovery of data for servicing.
Device as compact or remote version. 2-line backlit display with push buttons. HART, PROFIBUS PA, FOUNDATION Fieldbus.
Nominal diameter range2 path version: DN80 to 300 (3 to12")
3 path version: DN25 to 50 (1 to 2")
4 path version: DN 80 to 300 (3 to 12")

Wetted materialsSensor:
A351-CF3M (DN25 to 100)
1.4404/TP316/TP316L or A106 GrB (DN150 to 300)
Transducer: 1.4404/316/316L
Flanges : 1.4404/316/316L or A105/1.0432


Measured variablesVolume flow, calculated mass flow, sound velocity, flow velocity, signal strength
Max. measurement errorVolume flow (standard):
- ±0.5 % o.r. for 0.5 to 10 m/s (1.6 to 33 ft/s)
Volume flow (option):
- ±0.3 % o.r. for 0.5 to 10 m/s (1.6 to 33 ft/s)
Measuring range0.5 to 10 m/s (1.6 to 33 ft/s)
Max. process pressurePN 40 / ASME Cl. 300 / JIS 20K
Medium temperature range-40 to +150 °C (-40 to +302 °F)
-40 to +200 °C (-40 to +392 °F) optional
Ambient temperature range-40 to +60 °C (-40 to +140 °F) compact
-40 to +80 °C (-40 to +176 °F) remote sensor
-40 to +60 °C (-40 to +140 °F) remote transmitter
Transmitter housing materialAlSi10Mg, coated
Degree of protectionIP67, type 4X enclosure
IP68 type 6P enclosure (option for remote)
Display/Operation2 lines backlit display with 3 push buttons
Outputs1x 4-20 mA HART
1x Pulse/frequency/switch output (passive)
InputsN/A
Digital communicationHART, Profibus PA, FOUNDATION Fieldbus
Power supply2 wire loop powered
Hazardous area approvalsATEX,FM, CSA, JPNm UK Ex
Other approvals and certificates3.1 material (wetted parts), calibration performed on accredited
calibration facilities (acc. to ISO/IEC 17025), CRN, AD2000
PED, EAC marking
Product safetyEAC marking
Metrological approvals and certificatescalibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Pressure approvals and certificatesCRN, PED, AD2000
Material certificates3.1 material (wetted parts)
Returns Policy

Related Products

Top