Endress+Hauser Micropilot FMR62B – 80 GHz radar sensor

Write a Review

REQUEST A QUOTE

Endress+Hauser Micropilot FMR62B – 80 GHz radar sensor
Accuracy+/-1 mm (0.04 in)
Process temperature-196 … +450 °C (-321 … +842 °F)
Process pressure / max. overpressure limitVacuum...160 bar (2320 psi)
Max. measurement distance80 m (262 ft)
Main wetted parts316L or PTFE (clad 316L flanges)
Measuring principleLevel radar
Characteristic / ApplicationPremium device for continuous non-contact
level measurements, where aggressive media
are used; For vessels with many internals
SpecialitiesHeartbeat Technology;
SIL 2/3 according to IEC 61508;
Bluetooth® commissioning;
Operation and maintenance with SmartBlue App;
HistoROM;
RFID TAG for easy identification;
Very narrow beam angle
Supply / Communication2-wire (HART/ PROFIBUS PA)
PROFINET with Ethernet-APL
Bluetooth® wireless technology
FrequencyW-Band (~80 GHz)
Accuracy+/-1 mm (0.04 in)
Ambient temperature-40 … +85 °C (-40 … +185 °F)
Process temperature-196 … +450 °C (-321 … +842 °F)
Process pressure / max. overpressure limitVacuum...160 bar (2320 psi)
Main wetted parts316L or PTFE (clad 316L flanges)
Process connectionFlange:
DN50 ... DN150
ASME 2" ... 8"
Max. measurement distance80 m (262 ft)
CommunicationHART/ PROFIBUS PA
PROFINET with Ethernet-APL
Bluetooth® wireless
technology and app (optional)
Certificates / ApprovalsIEC Ex, ATEX, CSA C/US, EAC Ex, JPN Ex, KC Ex,
INMETRO Ex, NEPSI Ex, UK Ex
Safety approvalsOverfill protection WHG;
SIL 2/ SIL 3
Design approvalsAcceptance test certificate 3.1, EN10204;
NACE MR0175 / ISO 15156;
NACE MR0103 / ISO 17945;
EU Food Contact Material (EC) 1935/2004;
US Food Contact Material FDA CFR 21;
Declaration of Conformity ASME B31.3;
Conformity to cGMP;
AD2000;
PMI test;
CRN approval;
Helium leak test;
Liquid penetrant test AD2000-HP5-3(PT);
Liquid penetrant testing ISO23277-1 (PT);
Liquid penetrant testing ASME VIII-1 (PT);
Welding documentation (according to ISO or ASME)
Marine approvalABS, LR, BV, RINA, DNV
OptionsDisplay,
Customized settings,
App 'SmartBlue' for commissioning and operation via Bluetooth
Gas-tight implementation;
Overvoltage protection;
PWIS free
Application limitsStrong turbulences and foam,
long stilling wells (>12 m) and bypass -> FMR5x,FMP5x
Returns Policy

Related Products

Top