Endress+Hauser iTHERM ModuLine TM111

Write a Review

REQUEST A QUOTE

Endress+Hauser iTHERM ModuLine TM111
AccuracyClass AA acc. to IEC 60751
Class A acc. to IEC 60751
Class B acc. to IEC 60751
Class special or standard acc. to ASTM E230
Class 1 or 2 acc. to IEC 60584-2
Response timet90 starting at < 1,5 s QuickSens
depending on configuration
Max. process pressure (static)depending on the configuration
Operating temperature rangePT100 TF StrongSens:
-50 °C ...500 °C
(-58 °F ...932 °F)
PT100 QuickSensTF:
-50 °C …200 °C
(-58 °F …392 °F)
PT100 WW:
-200 °C ...600 °C
(-328 °F ...1.112 °F)
PT100 TF:
-50 °C ...400 °C
(-58 °F ...752 °F)
Typ K:
max. 1.100 °C
(max. 2.012 °F)
Typ J:
max. 800 °C
(max. 1.472 °F)
Typ N:
max. 1.100 °C
(max. 2.012 °F)
Max. immersion length on requestup to 4.500,0 mm (177'')
Measuring principleResistance Temperature Detector
Characteristic / Applicationmetric style
universal range of application
suitable for hazardous areas
can be used with StrongSens, QuickSens insert
direct process contact



Thermowell / protection tubewithout, direct process contact
Insert / probemineral insulated (MI), flexible
Outer diameter protection tube / InsertInsert:
3,0 mm
6,0 mm

Max. immersion length on requestup to 4.500,0 mm (177'')
Material protection tube/ thermowellInsert Material:
316L (1.4404)
Alloy 600 (2.4816)
Pyrosil


Process connectionThread:
G1/4, G1/2"
NPT1/2", NPT3/4"
M18x1.5, M20x1.5
Cap-nut:
G1/2", G3/4"
Compression fitting, also spring load:
NPT1/2", G1/2"
Weld-in adapter cylindrical or spherical







Tip shapestraight
Surface roughness Ra< 1.6 μm (63.00 μin)
Operating temperature rangePT100 TF StrongSens:
-50 °C ...500 °C
(-58 °F ...932 °F)
PT100 QuickSensTF:
-50 °C …200 °C
(-58 °F …392 °F)
PT100 WW:
-200 °C ...600 °C
(-328 °F ...1.112 °F)
PT100 TF:
-50 °C ...400 °C
(-58 °F ...752 °F)
Typ K:
max. 1.100 °C
(max. 2.012 °F)
Typ J:
max. 800 °C
(max. 1.472 °F)
Typ N:
max. 1.100 °C
(max. 2.012 °F)







Max. process pressure (static)depending on the configuration
AccuracyClass AA acc. to IEC 60751
Class A acc. to IEC 60751
Class B acc. to IEC 60751
Class special or standard acc. to ASTM E230
Class 1 or 2 acc. to IEC 60584-2
Response timet90 starting at < 1,5 s QuickSens
depending on configuration
Integration head transmitteryes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS)
Ex - approvalsATEX
ATEX IECEx
NEPSI
IECEx
EAC Ex
CSA C/US
INMETRO
CertificationSIL (Transmitter)