Endress+Hauser Iron analyzer Liquiline System CA80FE

Write a Review

REQUEST A QUOTE

Endress+Hauser Iron analyzer Liquiline System CA80FE
Measurement range0.05 to 2 mg/l Fe
0.1 to 5 mg/l Fe
0.1 to 5 mg/l with dilution function to maximum 2.0 to 100 mg/l Fe
Process temperature4 to 40 °C (39.2 to 104 °F)
Process pressureAt atmospheric pressure, < 0.2 bar
Measuring principleColorimetric
CharacteristicProcess analyzer for iron in aqueous solutions
Comply with standard colorimetric measuring principle, ferrozine method
SizeHousing (open version):
793 x 530 x 417 mm
31.22 x 20.87 x 16.42 in
Housing (closed version):
793 x 530 x 463 mm
31.22 x 20.87 x 18.23 in
Housing with base:
1723 x 530 x 463 mm
67.83 x 20.87 x 18.23 in
DesignOpen design, cabinet and stand housing
High-Performance plastic ASA-PC, additional stand coated steel
Process temperature4 to 40 °C (39.2 to 104 °F)
Ambient temperature5 to 40 °C (41 to 104 °F)
Outdoor version: - 20 to 40 °C (-4 to 104 °F)
Process pressureAt atmospheric pressure, < 0.2 bar
Sample flow rateMin. 5 ml/min (0.17 fl.oz/min)
Consistency of the sampleSuspended solids content
Turbitity < 50 NTU, aqueous, homogenized
ApplicationMonitoring of the iron content of potable water, mineral water and source water for industrial processes
Power supply100 to 120 VAC / 200 to 240 VAC ± 10%
24 VDC ± 10%
50 ± 1 or 60 ± 1,2 Hz
Output / communication2x 0/4 to 20 mA
Webserver, Modbus, Ethernet/IP, Profibus DP
Input1 or 2 measuring channel
1 to 4 digital sensor inputs for sensors with Memosens protocol (optional)
Measurement range0.05 to 2 mg/l Fe
0.1 to 5 mg/l Fe
0.1 to 5 mg/l with dilution function to maximum 2.0 to 100 mg/l Fe
ConsumablesReagents and standard solutions CY80FE for the operation
Regular maintenance is done with the parts of the maintenance kit CAV800
Returns Policy

Related Products

Top