Endress+Hauser Hydrostatic Level measurement Deltapilot FMB70

Write a Review

REQUEST A QUOTE

Endress+Hauser Hydrostatic Level measurement Deltapilot FMB70
AccuracyStandard 0.1%
Optional 0.075%
Process temperature-10°C...+100°C
(+14°F...+212°F)
Pressure measuring range100 mbar...10 bar
(1.5 psi...150 psi)
Process pressure / max. overpressure limit40 bar (600 psi)
Main wetted partsAlloy C
316L
Max. measurement distance100 m (328 ft) H2O
Material process membrane316L, AlloyC,
Gold-Rhodium
Measuring cell100 mbar...10 bar
(1.5 psi...150 psi)
Measuring principleHydrostatic pressure
CharacteristicPressure transmitter for hydrostatic level measurement with flush mounted metallic Contite measuring cell:
Hermetically sealed, condensate-resistant, climatic-proofed and with lowest temperature influences
Compact version

Supply voltage4...20 mA HART
10,5...45 VDC (Non Ex):
Ex ia: 10,5...30 VDC
PROFIBUS PA:
9...32 VDC (Non Ex)
FOUNDATION Fieldbus:
9...32 VDC (Non Ex)



Reference AccuracyStandard 0.1%
Optional 0.075%
Long term stability0.05 % of URL/year
Process temperature-10°C...100°C
(14°F...212°F)
Ambient temperature-40°C...85°C
(-40°F...185°F)
Measuring cell100 mbar...10 bar
(1.5 psi...150 psi)
Smallest calibratable span25 mbar (3.63 psi)
Vacuum resistance0 mbar abs.
Max. overpressure limit40 bar (600 psi)
Process connectionThread:
G1 1/2, MNPT1/2
Flange:
DN40...DN100,
ASME 1 1/2"...4",
JIS 10K
Process connection hygienicTri-Clamp
DIN11851
DIN11864-1
NEUMO
Varivent N
SMS
DRD
Material process membrane316L, AlloyC,
Gold-Rhodium
Material gasketNone, measuring cell welded
Fill fluidInert oil,
Synthetic oil
Material housing316L, Die-cast aluminum
Communication4 ... 20 mA HART
PROFIBUS PA
FOUNDATION Fieldbus
Certificates / ApprovalsATEX, FM, CSA, CSA C/US, IEC Ex, INMETRO, NEPSI, EAC, UK Ex
Safety approvalsSIL
Design approvalsEN10204-3.1
NACE MR0175
MR0103
Hygienic approvals3A, EHEDG
CoC ASME-BPE
Marine approvalsGL, DNV
Drinking water approvalsNSF
SpecialitiesGold-rhodium coated membrane
Measuring principleHydrostatic
Characteristic / ApplicationCompact pressure transmitter for hydrostatic level measurement with flush mounted metallic Contite measuring cell:
Hermetically sealed, condensate-resistant and climatic-proofed
SpecialitiesDiagnostic functions
Supply / Communication4...20mA HART:
10,5...45V DC
Ex ia: 10,5...30V DC
PROFIBUS PA /
FOUNDATION Fieldbus:
9...32V DC (Non Ex)
AccuracyStandard 0.1%
Optional 0.075%
Long term stability0.05 % of URL/year
Ambient temperature-40°C...+85°C
(-40°F...+185°F)
Process temperature-10°C...+100°C
(+14°F...+212°F)
Process pressure / max. overpressure limit40 bar (600 psi)
Pressure measuring range100 mbar...10 bar
(1.5 psi...150 psi)
Main wetted partsAlloy C
316L
Process connectionThread:
G1 1/2, MNPT1/2
Flange:
DN40...DN100,
ASME 1 1/2"...4",
JIS 10K
Process connection hygienicTri-Clamp
DIN11851
DIN11864-1
NEUMO
Varivent N
SMS
DRD
Max. measurement distance100 m (328 ft) H2O
Communication4...20 mA HART
PROFIBUS PA
FOUNDATION Fieldbus
Certificates / ApprovalsATEX, FM, CSA, CSA C/US, IEC Ex, INMETRO, NEPSI, EAC
Safety approvalsSIL
Design approvalsEN10204-3.1
NACE MR0175, MR0103
Hygienic approvals3A, EHEDG
CoC ASME-BPE
Marine approvalGL/ DNV
Drinking water approvalsNSF
OptionsHistoROM/M-Dat
4-line digital display
SS- or Aluminiumhousing
Separate housing
Application limitsMeasuring cell: Contite,
condensate-proof, watertight,
metal welded
If pressurized, possibly use
differential pressure measurement
with two pressure
transmitters. Observe ratio
head pressure : hydrostatic
pressure