Endress+Hauser
Endress+Hauser Digital ORP sensor Memosens CPS62E
REQUEST A QUOTE
Measurement range | –1 500 to 1 500 mV |
---|---|
Process temperature | 0 to 100 °C (32 to 212 °F) 0 to 140 °C (32 to 284 °F) (140 °C (284 °F) or sterilization only) |
Process pressure | 0.8 to 7 bar (11.6 to 101.5 psi) absolute |
Measuring principle | Sensor ORP / Redox |
Application | Hygiene and sterile applications (sterilizable, autoclavable): • Fermenter • Biotechnology • Pharma |
Characteristic | Digital ORP electrode for hygienic production processes with ion trap for long-term stable reference |
Measurement range | –1 500 to 1 500 mV |
Measuring principle | Gel compact electrode with ceramic junction and ion trap Measuring element platinum cap |
Design | All shaft lengths with temperature sensor Advanced gel technology |
Material | Sensor shaft: Glass to suit process Metal lead: Ag/AgCl Open aperture: Ceramic junction, zirconium dioxide ORP measuring element: Platinum O-ring: FKM Process coupling: PPS fiber-glass reinforced Nameplate: Ceramic metal oxide |
Dimension | Diameter: 12 mm (0.46 inch) Shaft lengths: 120, 225 and 360 mm (4.68, 8.77 and 14.04 inch) |
Process temperature | 0 to 100 °C (32 to 212 °F) 0 to 140 °C (32 to 284 °F) (140 °C (284 °F) or sterilization only) |
Process pressure | 0.8 to 7 bar (11.6 to 101.5 psi) absolute |
Temperature sensor | NTC 30k |
Ex certification | With ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in hazardous areas Zone 0, Zone 1 and Zone 2 |
Connection | Inductive, digital connection head with Memosens 2.0 technology |
Ingress protection | IP68 |
Additional certifications | Additional certifications |
Returns Policy