Endress+Hauser
Endress+Hauser Digital ORP sensor Memosens CPS42E
REQUEST A QUOTE
Measurement range | –1 500 to 1 500 mV |
---|---|
Process temperature | –15 to 135 °C (5 to 275 °F) |
Process pressure | 0.8 to 11 bar (11.6 to 159.5 psi) (absolute) |
Measuring principle | Sensor ORP / Redox |
Application | Media with very low conductivity or a high proportion of organic solvents or alcohol: • Chemical industry • Organic chemicals • Power stations • Laboratory measurements |
Characteristic | Digital ORP electrode for process engineering with ceramic junction and KCl liquid electrolyte |
Measurement range | –1 500 to 1 500 mV |
Measuring principle | Liquid-KCl filling and ceramic junction |
Design | All shaft lengths with temperature sensor |
Material | Sensor shaft: Glass to suit process ORP measuring element: Platinum Metal lead: Ag/AgCl Open aperture: Ceramic junction, zirconium dioxide O-ring: FKM Process coupling: PPS fiber-glass reinforced Nameplate: Ceramic metal oxide |
Dimension | Diameter: 12 mm (0.47 inch) Shaft length: 120, 225, 360 and 425 mm (4.72, 8.86, 14.17 and 16.73 inch) |
Process temperature | –15 to 135 °C (5 to 275 °F) |
Process pressure | 0.8 to 11 bar (11.6 to 159.5 psi) (absolute) |
Temperature sensor | NTC 30k |
Ex certification | With ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in hazardous areas Zone 0, Zone 1 and Zone 2. |
Connection | Inductive, contactless connection head with Memosens 2.0 technology |
Ingress protection | IP68 |
Additional certifications | Additional certifications |
Returns Policy