Endress+Hauser
Endress+Hauser Digital free chlorine sensor Memosens CCS51E
REQUEST A QUOTE
Measurement range | Trace: 0 to 5 mg/l HOCl Standard: 0 to 20 mg/l HOCl High: 0 to 200 mg/l HOCl |
---|---|
Process temperature | 0 to 55 °C (32 to 130 °F), non-freezing |
Process pressure | Max. 1 bar (max. 14.5 psi) |
Measuring method | Closed, membrane covered measuring cell Reduction of free chlorine at the cathode |
Measuring principle | Free chlorine |
Application | Ensuring reliable disinfection in drinking water Process water Dosing disinfectant efficiently in pool water Detect the absence or presence of free chlorine in Utilities Guarantee food safety and provide hygienic packaging and bottling |
Characteristic | Amperometric measurement of dissolved free chlorine |
Measurement range | Trace: 0 to 5 mg/l HOCl Standard: 0 to 20 mg/l HOCl High: 0 to 200 mg/l HOCl |
Measuring method | Closed, membrane covered measuring cell Reduction of free chlorine at the cathode |
Design | Closed amperometric 2-electrode measuring cell with PVDF membrane |
Material | Sensor shaft: POM Membrane: PVDF Membrane cap: PVDF Sealing ring: FKM |
Dimension | Diameter: 25 mm (0.98 inch) Length: 161 mm (6.34 inch) |
Process temperature | 0 to 55 °C (32 to 130 °F), non-freezing |
Process pressure | Max. 1 bar (max. 14.5 psi) |
Temperature sensor | 10k NTC integrated (Memosens) |
Connection | Inductive, digital connection head with Memosens |