Endress+Hauser Digital combined pH/ORP sensor Memosens CPS16E

Write a Review

REQUEST A QUOTE

Endress+Hauser Digital combined pH/ORP sensor Memosens CPS16E
Measurement rangeORP: –1 500 … 1 500 mV
Application: A
• pH: 1 to 12
Application B
• pH: 0 to 14
Process temperatureApplication A: –15 to 80 °C (5 to 176 °F)
Application B: 0 to 135 °C (32 to 275 °F)
Process pressure0.8 to 17 bar (11.6 to 246,5 psi) absolute
Measuring principlePotentiometric
ApplicationLong-term monitoring and limit control in processes with stable process conditions
• Chemical processes
• Pulp and paper industry
• Waste incinerators and gas scrubbers
• Mining
• Wastewater
CharacteristicCombination of pH and ORP electrode for standard applications in process technology and environmental engineering
Poison-resistant reference with ion trap
Measurement rangeORP: –1 500 to 1 500 mV
Application A
• pH: 1 to 12
Application B
• pH: 0 to 14
Measuring principleGel compact electrode with PTFE ring-junction and with ion trap
Pt-disc as additional ORP element
rH measurement and control of the reference impedance
DesignAll shaft lengths with temperature sensor
Advanced gel technology
MaterialSensor shaft: Glass to suit process
pH membrane glass: Type A, B
Metal lead: Ag/AgCl
Open aperture: PFTE junction, sterilizable
ORP measuring element: Platinum
O-ring: FKM
Process coupling: PPS fiber-glass reinforced
Nameplate: Ceramic metal oxide
DimensionDiameter: 12 mm (0.47 inch)
Shaft lengths: 120, 225 and 360 mm
(4.72, 8.86, 14.17 and 16.73 inch)
Process temperatureApplication A: –15 to 80 °C (5 to 176 °F)
Application B: 0 to 135 °C (32 to 275 °F)
Process pressure0.8 to 17 bar (11.6 to 246,5 psi) absolute
Temperature sensorNTC 30k
Ex certificationWith ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in
hazardous areas Zone 0, Zone 1 and Zone 2
ConnectionInductive, digital connection head with Memosens 2.0 technology
Ingress protectionIP68
Measuring principleSensor ORP / Redox
ApplicationLong-term monitoring and limit control in processes with stable process conditions
Chemical processes
• Pulp and paper industry
• Waste incinerators and gas scrubbers
• Mining
• Wastewater
CharacteristicCombination of pH and ORP electrode for standard applications in process technology and environmental engineering
Poison-resistant reference with ion trap
Measurement rangeORP: –1 500 … 1 500 mV
Application: A
• pH: 1 to 12
Application B
• pH: 0 to 14
Measuring principleGel compact electrode with PTFE ring-junction with ion trap
Pt-disc as additional ORP element
rH measurement and control of the reference impedance
DesignAll shaft lengths with temperature sensor
Advanced gel technology
MaterialSensor shaft: Glass to suit process
pH membrane glass: Type A, B
Metal lead: Ag/AgCl
Open aperture: PFTE junction, sterilizable
ORP measuring element: Platinum
O-ring: FKM
Process coupling: PPS fiber-glass reinforced
Nameplate: Ceramic metal oxide
DimensionDiameter: 12 mm (0.47 inch)
Shaft lengths: 120, 225 and 360 mm
(4.72, 8.86, 14.17 and 16.73 inch)
Process temperatureApplication A: –15 to 80 °C (5 to 176 °F)
Application B: 0 to 135 °C (32 to 275 °F)
Process pressure0.8 to 17 bar (11.6 to 246,5 psi) absolute
Temperature sensorNTK 30k
Ex certificationWith ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in
hazardous areas Zone 0, Zone 1 and Zone 2
ConnectionInductive, digital connection head with Memosens 2.0 technology
Ingress protectionIP68