Endress+Hauser Deltabar PMD55B - differential pressure transmitter

Write a Review

REQUEST A QUOTE

Endress+Hauser Deltabar PMD55B - differential pressure transmitter
AccuracyStandard:
up to 0.075 %
Platinum:
up to 0.055 %
Process temperature-40°C...+110°C (-40°F...+230°F)
Pressure measuring range30 mbar.... 40 bar (0.45 psi... 600 psi)
Main wetted parts316L, AlloyC
Material process membrane316L, AlloyC
Measuring cell30 mbar...40 bar
(0.45 psi...600 psi)
Measuring principleDifferential pressure
CharacteristicTransmitter with metal membrane for differential pressure monitoring in liquids or gases.
Intuitive and clear wizard for commissioning
Supply voltageEx d, Ex e, non-Ex:
10.5...35V DC
Ex i: 10.5...30V DC

Reference AccuracyStandard:
up to 0.075 %
Platinum:
up to 0.055 %
Long term stability0.05 % of URL/ year
0.13 % of URL/ 5 years
0.23 % of URL/ 10 years
Process temperature-40°C...+110°C
(-40°F...+230°F)
Ambient temperature-40°C...+85°C
(-40°F...+185°F)
Measuring cell30 mbar...40 bar
(0.45 psi...600 psi)
Smallest calibratable span0.3 mbar (0.0045 psi)
Vacuum resistance25 mbar (0.0375 psi)
Max. Turn down100:1
Max. overpressure limit250 bar (3750 psi)
Process connectionNPT1/4-18,
RC1/4"
Material process membrane316L, AlloyC
Material gasketPTFE, FKM, EPDM, FFKM
Fill fluidSilicone oil,
Inert oil
Material housingAluminium
Communication4...20 mA HART
PROFINET with Ethernet-APL
Bluetooth® wireless
technology and app (optional)
Certificates / ApprovalsATEX, CSA, IECEx, JPN Ex, INMETRO, NEPSI, UK Ex
Safety approvalsSIL
Design approvalsEN10204-3.1,
NACE MR0175,
NACE MR0103
Drinking water approvalsNSF, KTW
SpecialitiesBluetooth® Operation and maintenance SmartBlue App,
RFID TAG for easy identification,
Plug and play functionalities
Measuring principleDifferential pressure
Characteristic / ApplicationTransmitter with metal membrane for differential pressure monitoring in liquids or gases.
Intuitive and clear wizard for commissioning
SpecialitiesBluetooth® Operation and maintenance with SmartBlue App,
RFID TAG for easy identification,
Plug and play functionalities
Supply / CommunicationEx d, Ex e, non-Ex::
10.5...35V DC
Ex i: 10.5...30V DC
AccuracyStandard:
up to 0.075 %
Platinum:
up to 0.055 %
Long term stability0.05 % of URL/ year
0.13 % of URL/ 5 years
0.23 % of URL/ 10 years
Ambient temperature-40°C...+85°C (-40°F...+185°F)
Process temperature-40°C...+110°C (-40°F...+230°F)
Process pressure / max. overpressure limit250 bar (3750 psi)
Pressure measuring range30 mbar.... 40 bar (0.45 psi... 600 psi)
Main wetted parts316L, AlloyC
Process connectionNPT 1/4-18, RC1/4"
Communication4...20 mA HART
PROFINET with Ethernet-APL
Bluetooth® wireless
technology and app (optional)
Certificates / ApprovalsATEX, CSA, IECEx, JPN Ex, INMETRO, NEPSI, UK Ex
Safety approvalsSIL
Design approvalsEN 10204-3.1,
NACE MR0175,
NACE MR0103,
AD2000
Drinking water approvalsNSF, KTW
Application limitsNone
Measuring principleDifferential pressure
Product headlineTransmitter with metal membrane for differential pressure monitoring in liquids or gases.
Intuitive and clear wizard for commissioning
Sensor featuresFill fluid:
Inert oil,
Silicone oil
Transmitter featuresBluetooth® Operation and maintenance SmartBlue App,
RFID TAG for easy identification,
Plug and play functionalities
Wetted materials316L, Alloy
Measuring range30 mbar...40 bar
(0.45 psi...600 psi)
Medium temperature range-40°C...+110°C
(-40°F...+230°F)
Ambient temperature range-40°C...+85°C
(-40°F...+185°F)
Transmitter housing materialAluminium
Display/OperationConfiguration via local display and operating tools possible
Remote display available
Outputs4...20 mA HART
Digital communicationHART
Hazardous area approvalsATEX, CSA, IECEx
Functional safetySIL
Material certificatesEN10204-3.1,
NACE MR0175,
NACE MR0103
Measuring principleDifferential pressure
Measuring principleDifferential pressure
Product headlineTransmitter with metal membrane for differential pressure monitoring in liquids or gases.
Intuitive and clear wizard for Commissioning
Sensor featuresFill fluid:
Inert oil,
Silicone oil
Transmitter featuresBluetooth® Operation and maintenance SmartBlue App,
RFID TAG for easy identification,
Plug and play functionalities
Wetted materials316L, Alloy
Measuring range30 mbar...40 bar
(0.45 psi...600 psi)
Medium temperature range-40°C...+110°C
(-40°F...+230°F)
Ambient temperature range-40°C...+85°C
(-40°F...+185°F)
Transmitter housing materialAluminium
Display/OperationConfiguration via local display and operating tools possible
Remote display available
Outputs4...20 mA HART
Digital communicationHART
Hazardous area approvalsATEX, CSA, IECEx, JPN Ex, INMETRO, NEPSI, UK Ex
Functional safetySIL
Material certificatesEN10204-3.1,
NACE MR0175,
NACE MR0103
AD2000
Returns Policy