Endress+Hauser Chromate analyzer Liquiline System CA80CR

Write a Review

REQUEST A QUOTE

Endress+Hauser Chromate analyzer Liquiline System CA80CR
Measurement range0,03 to 2.5 mg/l Cr(VI)
0.2 to 5.0 mg/l Cr(VI)
0.2 to 5.0 mg/l with dilution function to maximum 10 to 250 mg/l Cr(VI)
Process temperature4 to 40 °C (39 to 104 °F)
Process pressureAt atmospheric pressure, < 0.2 bar absolute
Measuring principleColorimetric
CharacteristicAnalyzer for Chromate (Chrome VI) in aqueous solutions
SizeHousing (open version):
793 x 530 x 417 mm
31.22 x 20.87 x 16.42 in
Housing (closed version):
793 x 530 x 463 mm
31.22 x 20.87 x 18.23 in
Housing with base:
1723 x 530 x 463 mm
67.83 x 20.87 x 18.23 in

DesignOpen design, cabinet and stand housing
High-Performance plastic ASA-PC, additional stand coated steel
Process temperature4 to 40 °C (39 to 104 °F)
Ambient temperature5 to 40 °C (41 to 104 °F)
Outdoor version: - 20 to 40 °C (-4 to 104 °F)
Process pressureAt atmospheric pressure, < 0.2 bar absolute
Sample flow rateMin. 5 ml/min (0.17 fl.oz/min)
Consistency of the sampleSuspended solids content
Turbitity < 50 NTU, aqueous, homogenized
SpecialsEasy upgrade to measuring station with up to four digital Memosens sensors
Automatic calibration and cleaning
User-configurable measuring, cleaning and calibration intervals
Optional cooling module for longer reagent shelf life
2 channel version optional
User-definable measuring ranges
Modular design for easily extensible functionality
Digital communication for remote access
ApplicationLimit value monitoring of Chromate in potable water and waste water
Common applications are industrial waste water Monitoring for tanneries and the galvanic industry
Power supply100 to 120 VAC / 200 to 240 VAC ± 10%
24 VDC ± 10%
50 ± 1 or 60 ± 1,2 Hz
Output / communication2x 0/4 to 20 mA
Modbus RS485/TCP (optional)
Ethernet (optional)
Alarmrelay
Input1 or 2 measuring channels
1 to 4 digital sensor inputs for sensors with Memosens protocol (optional)
Measurement range0,03 to 2.5 mg/l Cr(VI)
0.2 to 5.0 mg/l Cr(VI)
0.2 to 5.0 mg/l with dilution function to maximum 10 to 250 mg/l Cr(VI)
ConsumablesReagents and standard solutions CY80CR as well as cleaner CY800 are necessary for the operation
Regular maintenance is done with the parts of the maintenance kit CAV800