Endress+Hauser
Endress+Hauser Aluminum analyzer Liquiline System CA80AL
REQUEST A QUOTE
Measurement range | 15 to 1000 µg/l Al 15 to 1000 µg/l with dilution function to maximum 300 to 20 000 µg/l Al |
---|---|
Process temperature | 4 ... 40 °C (39.2 ... 104 °F) |
Process pressure | at atmospheric pressure, < 0.2 bar |
Measuring principle | Colorimetric |
Characteristic | Process analyzer for aluminum in aqueous solutions Comply with standard colorimetric measuring principle - pyrocatechol violet methot |
Size | Housing (open version): 793 x 530 x 417 mm 31.22 x 20.87 x 16.42 in Housing (closed version): 793 x 530 x 463 mm 31.22 x 20.87 x 18.23 in Housing with base: 1723 x 530 x 463 mm 67.83 x 20.87 x 18.23 in |
Design | Open design, cabinet and stand housing High-Performance plastic ASA-PC, additional stand coated steel |
Process temperature | 4 ... 40 °C (39.2 ... 104 °F) |
Ambient temperature | 5 ... 40C ( 41 ... 104F) Outdoor version: - 20 to 40 °C (-4 to 104 °F) |
Process pressure | at atmospheric pressure, < 0.2 bar |
Sample flow rate | min. 5ml/min (0,17 fl.oz/min) |
Consistency of the sample | low solid content (< 50NTU), aqueous homogeneous |
Application | Monitoring of the aluminum content of potable water, mineral water and industrial waste water |
Power supply | 100 ... 120 VAC / 200 ... 240 VAC ± 10% 24 VDC ± 10% 50 ± 1 oder 60 ± 1,2 Hz |
Output / communication | 2x 0/4 ... 20 mA Webserver, Modbus, Ethernet/IP, Profibus DP |
Input | 1 or 2 measuring channel 1 ... 4 digital sensor inputs for sensors with Memosens protocol (optional) |
Measurement range | 15 to 1000 µg/l Al 15 to 1000 µg/l with dilution function to maximum 300 to 20 000 µg/l Al |
Consumables | Reagents and standard solutions CY80AL as well as cleaner CY800 are necessary for the operation. Regular maintenance is done with the parts of the maintenance kit CAV800. |