Endress+Hauser Aluminum analyzer Liquiline System CA80AL
Measurement range15 to 1000 µg/l Al
15 to 1000 µg/l with dilution function to maximum 300 to 20 000 µg/l Al
Process temperature4 ... 40 °C (39.2 ... 104 °F)
Process pressureat atmospheric pressure, < 0.2 bar
Measuring principleColorimetric
CharacteristicProcess analyzer for aluminum in aqueous solutions
Comply with standard colorimetric measuring principle - pyrocatechol violet methot
SizeHousing (open version):
793 x 530 x 417 mm
31.22 x 20.87 x 16.42 in
Housing (closed version):
793 x 530 x 463 mm
31.22 x 20.87 x 18.23 in
Housing with base:
1723 x 530 x 463 mm
67.83 x 20.87 x 18.23 in
DesignOpen design, cabinet and stand housing
High-Performance plastic ASA-PC, additional stand coated steel
Process temperature4 ... 40 °C (39.2 ... 104 °F)
Ambient temperature5 ... 40C ( 41 ... 104F)
Outdoor version: - 20 to 40 °C (-4 to 104 °F)
Process pressureat atmospheric pressure, < 0.2 bar
Sample flow ratemin. 5ml/min (0,17 fl.oz/min)
Consistency of the samplelow solid content (< 50NTU), aqueous homogeneous
ApplicationMonitoring of the aluminum content of potable water, mineral water and industrial waste water
Power supply100 ... 120 VAC / 200 ... 240 VAC ± 10%
24 VDC ± 10%
50 ± 1 oder 60 ± 1,2 Hz
Output / communication2x 0/4 ... 20 mA
Webserver, Modbus, Ethernet/IP, Profibus DP
Input1 or 2 measuring channel
1 ... 4 digital sensor inputs for sensors with Memosens protocol (optional)
Measurement range15 to 1000 µg/l Al
15 to 1000 µg/l with dilution function to maximum 300 to 20 000 µg/l Al
ConsumablesReagents and standard solutions CY80AL as well as cleaner CY800 are necessary for the operation. Regular maintenance is done with the parts of the maintenance kit CAV800.