Endress+Hauser
Endress+Hauser 8FE LPGmass Coriolis flowmeter
REQUEST A QUOTE
Max. measurement error | Mass flow (liquid): ±0.2 % Volume flow (liquid): ±0.3 % |
---|---|
Measuring range | 0 to 45 000 kg/h (0 to 1650 lb/min) |
Medium temperature range | –40 to +125 °C (–40 to +257 °F) |
Max. process pressure | PN 40, Class 300, 63K |
Wetted materials | Measuring tube: 1.4539 (904L) Connection: 1.4404 (316/316L) |
Measuring principle | Coriolis |
Product headline | The refueling and distribution application flowmeter with easy sytem integration. Accurate measurement of liquefied petroleum gas in refueling and distribution applications. |
Sensor features | Excellent operational safety – reliable under extreme ambient conditions. Fewer process measuring points – multivariable measurement (flow, density, temperature). Space‐saving installation – no in/outlet run needs. Flow rates up to 45 000 kg/h (1654 lb/min). Volume flow calculation according to API table 53. |
Transmitter features | Easy operation – reduced to application needs. Fast commissioning – pre‐configured devices. Automatic recovery of data for servicing. Robust, ultra‐compact transmitter housing. Pulse output and Modbus RS485. |
Nominal diameter range | DN 8 to 40 (³⁄₈ to 1½") |
Wetted materials | Measuring tube: 1.4539 (904L) Connection: 1.4404 (316/316L) |
Measured variables | Mass flow, density, temperature, volume flow, corrected volume flow |
Max. measurement error | Mass flow (liquid): ±0.2 % Volume flow (liquid): ±0.3 % |
Measuring range | 0 to 45 000 kg/h (0 to 1650 lb/min) |
Max. process pressure | PN 40, Class 300, 63K |
Medium temperature range | –40 to +125 °C (–40 to +257 °F) |
Ambient temperature range | –40 to +60 °C (–40 to +140 °F) |
Sensor housing material | 1.4301 (304), corrosion resistant |
Transmitter housing material | Powder‐coated die‐cast aluminium |
Degree of protection | IP67, type 4X enclosure |
Display/Operation | No local operation Configuration via operating tools possible |
Outputs | Pulse/frequency/switch output (passive), phase‐shifted pulse |
Inputs | None |
Digital communication | Modbus RS485 |
Power supply | DC 10 to 30 V AC 20 to 28 V |
Hazardous area approvals | ATEX, IECEx, NEC/CEC, FM, CSA, NEPSI, UL |
Other approvals and certificates | 3.1 material, calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), custody transfer PED, CRN |
Metrological approvals and certificates | Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), custody transfer |
Pressure approvals and certificates | PED, CRN |
Material certificates | 3.1 material |
Returns Policy