Endress+Hauser
Endress+Hauser 8BE Dosimass Coriolis flowmeter
REQUEST A QUOTE
Max. measurement error | Mass flow (liquid): ±0.15 % Density (liquid): ±0.0005 g/cm³ |
---|---|
Measuring range | 0 to 18 000 kg/h (0 to 660 lb/min) |
Medium temperature range | –40 to +125 °C (–40 to +257 °F) 150 °C allowed for 1 h |
Max. process pressure | PN 40 |
Wetted materials | Measuring tube: 1.4539 (904L) Connection: 1.4404 (316/316L) |
Measuring principle | Coriolis |
Product headline | The compact sensor with an ultra-compact transmitter. Measurement of liquids with the most diverse properties in continuous process control & batching applications. |
Sensor features | High process safety – high measuring accuracy for different media in shortest filling time. Fewer process measuring points – multivariable measurement (flow, density, temperature). Space‐saving installation – no in-/outlet run needs. Nominal diameter: DN 8 to 25 (³⁄₈ to 1"). Various hygienic process connections, 3-A conform. CIP, SIP cleanable sensor. |
Transmitter features | Versatile and time-saving wiring – plug connector. Fast commissioning – pre-configured devices. Automatic recovery of data for servicing. Robust, ultra‐compact transmitter housing. Pulse/frequency/switch output, Modbus RS485. Excellent and easily cleanable transmitter. |
Nominal diameter range | DN 8 to 25 (³⁄₈ to 1") |
Wetted materials | Measuring tube: 1.4539 (904L) Connection: 1.4404 (316/316L) |
Measured variables | Mass flow, density, temperature, volume flow |
Max. measurement error | Mass flow (liquid): ±0.15 % Density (liquid): ±0.0005 g/cm³ |
Measuring range | 0 to 18 000 kg/h (0 to 660 lb/min) |
Max. process pressure | PN 40 |
Medium temperature range | –40 to +125 °C (–40 to +257 °F) 150 °C allowed for 1 h |
Ambient temperature range | –20 to +60 °C (–4 to +140 °F) |
Sensor housing material | 1.4301 (304), corrosion resistant |
Transmitter housing material | 1.4308 (304) |
Degree of protection | IP67, type 4X enclosure |
Display/Operation | No local operation Configuration via operating tools possible |
Outputs | Pulse/frequency/switch output (passive) |
Inputs | None |
Digital communication | Modbus RS485 |
Power supply | DC 20 to 30 V |
Hazardous area approvals | ATEX, IECEx, cCSAus |
Other approvals and certificates | 3.1 material, calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR 3-A |
Product safety | CE |
Returns Policy