Endress+Hauser 71488114 iTHERM TrustSens self-calibration Monitoring

Write a Review

REQUEST A QUOTE

Endress+Hauser 71488114 iTHERM TrustSens self-calibration Monitoring
  • All required components in one bundle, nothing is forgotten

  • Tamper-proof data storage according to FDA 21 CFR Part 11

  • Predictive maintenance

  • Calibration with timestamp

  • Higher product safety

  • With automated iTHERM TrustSens calibration monitoring, Endress+Hauser offers optimized process verification at minimum effort to users in the life sciences and food & beverage industries.
    The product bundle consisting of iTHERM TrustSens self-calibrating temperature sensors and the Advanced Data Manager Memograph M RSG45 offers FDA-compliant process verification, automated, audit-proof documentation and enables predictive maintenance.
    Reliable & easy:
    • Cost reduction by saving calibration costs and reducing calibration cycles

    100% compliance and audit-proof documentation:
    • Fully traceable calibration with automated documentation

    Risk reduction and elimination of undetected failures:
    • Higher process transparency

    Measuring principleData manager
    Measuring principleEnergymanager
    FunctionSecure data recording and signal analysis
    visualization
    integrated web server for device set-up
    HART(R) transparency, RSG45 as a HART modem
    E-Mail notification (encrypted mails)
    100 ms scan rate, high speed storage cycle 100 ms
    modular design for simple expansion
    data readout and processing via Field Data Manager Software
    OPC server for indication of instantaneous values (optional)
    Input20 analog (max) (U, I, TC, RTD, frequency, pulse)
    20 HART (max) (HART, 4…20mA)
    14 digital (max) (control, event and counting inputs)
    Output12 relays (max)
    2 analog (max) (I, pulse)
    1 transmitter power supply
    InterfaceRS232/485
    Ethernet
    USB
    Modbus RTU/TCP Slave
    Modbus RTU/TCP Master
    Profibus DP
    PROFINET Device
    EtherNet/IP Adapter
    Storage/Recordinginternal memory
    SD card
    USB flash drive
    DisplayTFT 7" colour graphic display
    TFT 7" colour graphic display with touch panel
    Housingpanel mounting with zink diecast bezel
    panel mounting with stainless steel bezel
    desktop housing
    field housing
    Power supply100...230 V AC +/-10%
    24V (-10%; +15%) AC/DC
    CertificatesCE
    EAC
    UL (US und CA)
    ATEX II2G Ex px IIC Gb
    ATEX II2D Ex pD IIIC Db
    UL listed
    PROFINET certified
    Software packageBatch software
    Waste water software
    Tele-alarm software
    Energy software (water + steam)
    Waste water + storm overflow + Tele Alarm
    Tele-alarm software + energy software (water + steam)
    Measuring principleEnergy manager
    Measuring principleEnergymanager
    FunctionMass flow and energy calculation for Water and Steam.
    CalculationsMass/heat quantity
    heat quantity difference
    Number of applicationsNot defined
    Data storageyes
    Calculation standardsIAPWS-97
    CommunicationRS232/485
    Ethernet
    USB
    Modbus RTU/TCP Slave
    Modbus RTU/TCP Master
    Profibus DP
    PROFINET Device
    EtherNet/IP Adapter
    Power supplyNot defined
    Loop power supply24 V DC +/-15%
    Protection classIP65 (front-panel)
    Input0...20 universal
    HART
    6…14 digital
    Output2 analog
    Dimensions (WxHxD)195,2 x 149,2 x 227,1 mm (7,69" x 5,87" x 8,94")
    OperationNavigator on-site, web server and via FieldCare/DeviceCare
    Display7“ TFT
    Software functionsInternal calculation for mass flow and energy
    CertificatesCE
    EAC
    UL (US und CA)
    ATEX II2G Ex px IIC Gb
    ATEX II2D Ex pD IIIC Db
    UL listed
    PROFINET certified
    Measuring principleResistance Temperature Detector
    Characteristic / Applicationself-calibrating
    metric style
    compact temperature probe
    hygienic design
    life sciences, food & beverage
    Thermowell / protection tubewithout
    incl. thermowell
    incl. Tee-/elbow thermowell
    Insert / probewithout exchangeable insert
    Outer diameter protection tube / Insert6.0 mm (0.24'')
    9.0 mm (0.35'')
    12.7 mm (0.5")
    Max. immersion length on requestup to 900.00 mm (35.4'')
    Material protection tube/ thermowell1.4435 Delta Ferrite < 1%
    316L
    Process connectioncompression fitting
    weld in adapter
    clamp connections acc. to ISO 2852
    screwed pipe joint acc. to DIN 11851
    aseptic screwed pipe joint acc. to DIN 11864-1
    metallic sealing system
    thread acc. to ISO 228 for Liquiphant adaptor
    APV Inline
    Varivent
    Ingold
    SMS 1147
    Neumo Biocontrol
    TT411 elbow thermowell DIN 11865
    Tip shapestraight
    reduced
    Surface roughness Ra0.76 μm (29.92 μin)
    0.38 μm (14.96 μin)
    0.38 μm (14.96 μin) electropolished
    Operating temperature rangePt100:
    -40 °C to 160 °C (-40 °F to 320 °F),
    optional up to 190 °C (374 °F)
    reference point for automated calibration
    Max. process pressure (static)at 20 °C: 40 bar (580 psi)
    Response timet50 = 2.5 s
    t90 = 5.4 s
    Integration head transmitterno (4 to 20 mA signal, HART)
    Ex - approvalsCSA C/US IS, I/1+2/A-F
    CSA C/US General Purpose
    UK II1/2G Ex ia IIC T6 Ga/Gb, II1/2D Ex ia IIIC Da/Db
    ATEX IECEx II1/2G Ex ia IIC T6 Ga/Gb, II1/2D Ex ia IIIC T6 Da/Db