Endress+Hauser
Endress+Hauser 71488114 iTHERM TrustSens self-calibration Monitoring
REQUEST A QUOTE
With automated iTHERM TrustSens calibration monitoring, Endress+Hauser offers optimized process verification at minimum effort to users in the life sciences and food & beverage industries. The product bundle consisting of iTHERM TrustSens self-calibrating temperature sensors and the Advanced Data Manager Memograph M RSG45 offers FDA-compliant process verification, automated, audit-proof documentation and enables predictive maintenance. | |
Reliable & easy: |
|
---|---|
100% compliance and audit-proof documentation: |
|
Risk reduction and elimination of undetected failures: |
|
Measuring principle | Data manager |
Measuring principle | Energymanager |
Function | Secure data recording and signal analysis visualization integrated web server for device set-up HART(R) transparency, RSG45 as a HART modem E-Mail notification (encrypted mails) 100 ms scan rate, high speed storage cycle 100 ms modular design for simple expansion data readout and processing via Field Data Manager Software OPC server for indication of instantaneous values (optional) |
Input | 20 analog (max) (U, I, TC, RTD, frequency, pulse) 20 HART (max) (HART, 4…20mA) 14 digital (max) (control, event and counting inputs) |
Output | 12 relays (max) 2 analog (max) (I, pulse) 1 transmitter power supply |
Interface | RS232/485 Ethernet USB Modbus RTU/TCP Slave Modbus RTU/TCP Master Profibus DP PROFINET Device EtherNet/IP Adapter |
Storage/Recording | internal memory SD card USB flash drive |
Display | TFT 7" colour graphic display TFT 7" colour graphic display with touch panel |
Housing | panel mounting with zink diecast bezel panel mounting with stainless steel bezel desktop housing field housing |
Power supply | 100...230 V AC +/-10% 24V (-10%; +15%) AC/DC |
Certificates | CE EAC UL (US und CA) ATEX II2G Ex px IIC Gb ATEX II2D Ex pD IIIC Db UL listed PROFINET certified |
Software package | Batch software Waste water software Tele-alarm software Energy software (water + steam) Waste water + storm overflow + Tele Alarm Tele-alarm software + energy software (water + steam) |
Measuring principle | Energy manager |
Measuring principle | Energymanager |
Function | Mass flow and energy calculation for Water and Steam. |
Calculations | Mass/heat quantity heat quantity difference |
Number of applications | Not defined |
Data storage | yes |
Calculation standards | IAPWS-97 |
Communication | RS232/485 Ethernet USB Modbus RTU/TCP Slave Modbus RTU/TCP Master Profibus DP PROFINET Device EtherNet/IP Adapter |
Power supply | Not defined |
Loop power supply | 24 V DC +/-15% |
Protection class | IP65 (front-panel) |
Input | 0...20 universal HART 6…14 digital |
Output | 2 analog |
Dimensions (WxHxD) | 195,2 x 149,2 x 227,1 mm (7,69" x 5,87" x 8,94") |
Operation | Navigator on-site, web server and via FieldCare/DeviceCare |
Display | 7“ TFT |
Software functions | Internal calculation for mass flow and energy |
Certificates | CE EAC UL (US und CA) ATEX II2G Ex px IIC Gb ATEX II2D Ex pD IIIC Db UL listed PROFINET certified |
Measuring principle | Resistance Temperature Detector |
Characteristic / Application | self-calibrating metric style compact temperature probe hygienic design life sciences, food & beverage |
Thermowell / protection tube | without incl. thermowell incl. Tee-/elbow thermowell |
Insert / probe | without exchangeable insert |
Outer diameter protection tube / Insert | 6.0 mm (0.24'') 9.0 mm (0.35'') 12.7 mm (0.5") |
Max. immersion length on request | up to 900.00 mm (35.4'') |
Material protection tube/ thermowell | 1.4435 Delta Ferrite < 1% 316L |
Process connection | compression fitting weld in adapter clamp connections acc. to ISO 2852 screwed pipe joint acc. to DIN 11851 aseptic screwed pipe joint acc. to DIN 11864-1 metallic sealing system thread acc. to ISO 228 for Liquiphant adaptor APV Inline Varivent Ingold SMS 1147 Neumo Biocontrol TT411 elbow thermowell DIN 11865 |
Tip shape | straight reduced |
Surface roughness Ra | 0.76 μm (29.92 μin) 0.38 μm (14.96 μin) 0.38 μm (14.96 μin) electropolished |
Operating temperature range | Pt100: -40 °C to 160 °C (-40 °F to 320 °F), optional up to 190 °C (374 °F) reference point for automated calibration |
Max. process pressure (static) | at 20 °C: 40 bar (580 psi) |
Response time | t50 = 2.5 s t90 = 5.4 s |
Integration head transmitter | no (4 to 20 mA signal, HART) |
Ex - approvals | CSA C/US IS, I/1+2/A-F CSA C/US General Purpose UK II1/2G Ex ia IIC T6 Ga/Gb, II1/2D Ex ia IIIC Da/Db ATEX IECEx II1/2G Ex ia IIC T6 Ga/Gb, II1/2D Ex ia IIIC T6 Da/Db |