Endress+Hauser 6BAB Proline t-mass B 150 thermal mass flowmeter

Write a Review

REQUEST A QUOTE

Endress+Hauser 6BAB Proline t-mass B 150 thermal mass flowmeter
Max. measurement error3 % o.r.
4 % o.r.
5 % o.f.s.
(depending on chosen option of ordering feature "Calibration flow")
Measuring range20 to 720 000 kg/h (45 to 1 587 600 lb/h)
20 to 1 080 000 kg/h (45 to 2 381 400 lb/h)
(for air, depending on chosen option of ordering feature "Calibration flow")
Medium temperature range–40 to +100 °C (–40 to +212 °F)
Max. process pressure20 bar g (290 psi g)
Wetted materialsTransducer: 1.4404 (316L)
Insertion tube: 1.4404 (316L); 1.4435 (316L)
Connection:
Compression fitting: 1.4404 (316L)
Sealing ring: EPDM; HNBR; 1.4401 (316)
Clamping ring: PEEK 450G
Measuring principleThermal
Product headlineThe flowmeter for cost-effective measurement and easy monitoring of utility gases.
Suitable for air, nitrogen, carbon dioxide and argon in circular piping or rectangular ducts.
Sensor featuresOptimal process monitoring – easy measurement even at low pressures and flow velocities. Cost‐effective measurement – easy installation, negligible pressure loss and maintenance‐free. Reliable flow trending – multivariable measurement.
Insertion version for nominal diameter DN 80 to 1500 (3 to 60"). Installation and removal of sensor without process interruption. Easy installation.
Transmitter featuresFast and efficient commissioning – guided operating menus. High plant availability – self‐diagnostics and error monitoring. Automatic recovery of data for servicing.
Device in compact version with DC 24 V power supply. 4-20 mA HART, pulse/frequency/switch output. Compact and robust transmitter.
Nominal diameter rangeDN 80 to 1500 (3 to 60")
Wetted materialsTransducer: 1.4404 (316L)
Insertion tube: 1.4404 (316L); 1.4435 (316L)
Connection:
Compression fitting: 1.4404 (316L)
Sealing ring: EPDM; HNBR; 1.4401 (316)
Clamping ring: PEEK 450G




Measured variablesMass flow, temperature, corrected volume flow, FAD volume flow
Max. measurement error3 % o.r.
4 % o.r.
5 % o.f.s.
(depending on chosen option of ordering feature "Calibration flow")

Measuring range20 to 720 000 kg/h (45 to 1 587 600 lb/h)
20 to 1 080 000 kg/h (45 to 2 381 400 lb/h)
(for air, depending on chosen option of ordering feature "Calibration flow")
Max. process pressure20 bar g (290 psi g)
Medium temperature range–40 to +100 °C (–40 to +212 °F)
Ambient temperature range–40 to +60 °C (–40 to +140 °F)
Transmitter housing materialAlSi10Mg, coated
Degree of protectionIP66/67, type 4X enclosure
Display/Operation4‐line display with push Buttons
Configuration via local display and operating tools possible
Outputs4‐20 mA HART (active)
Pulse/frequency/switch output (passive)
InputsStatus input
Digital communicationHART
Power supplyDC 18 to 30 V
Hazardous area approvalsATEX, IECEx, cCSAus
Metrological approvals and certificatesCalibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR
Pressure approvals and certificatesCRN
Returns Policy

Related Products