Endress+Hauser 6AAB Proline t-mass A 150 thermal mass flowmeter

Write a Review

REQUEST A QUOTE

Endress+Hauser 6AAB Proline t-mass A 150 thermal mass flowmeter
Max. measurement error3 % o.r.
4 % o.r.
5 % o.f.s.
(depending on chosen option of ordering feature "Calibration flow")
Measuring range0.5 to 910 kg/h (1.1 to 2002 lb/h)
0.5 to 1365 kg/h (1.1 to 3003 lb/h)
(for air, depending on chosen option of ordering feature "Calibration flow")
Medium temperature range–40 to +100 °C (–40 to +212 °F)
Max. process pressurePN 40, Class 300
Wetted materialsTransducer: 1.4404 (316L)
Insertion tube: 1.4404 (316L)
Measuring tube: 1.4404 (316L); 1.4435 (316L)
Connection: 1.4404 (F316/F316L); 1.4404 (316L); 1.4435 (316L)
Measuring principleThermal
Product headlineThe flowmeter for cost-effective measurement and easy monitoring of utility gases.
Suitable for air, nitrogen, carbon dioxide and argon in small line sizes.
Sensor featuresOptimal process monitoring – easy measurement even at low pressures and flow velocities. Cost-effective measurement – easy installation, negligible pressure loss and maintenance-free. Reliable flow trending – multivariable measurement.
Inline version: DN 15 to 50 (½ to 2"). Process pressure up to PN 40, Class 300. A variety of process connections available.
Transmitter featuresFast and efficient commissioning – guided operating menus. High plant availability – self‐diagnostics and error monitoring. Automatic recovery of data for servicing.
Device in compact version with DC 24 V power supply. 4-20 mA HART, pulse/frequency/switch output. Compact and robust transmitter.
Nominal diameter rangeDN 15 to 50 (½ to 2")
Wetted materialsTransducer: 1.4404 (316L)
Insertion tube: 1.4404 (316L)
Measuring tube: 1.4404 (316L); 1.4435 (316L)
Connection: 1.4404 (F316/F316L); 1.4404 (316L); 1.4435 (316L)


Measured variablesMass flow, temperature, corrected volume flow, FAD volume flow
Max. measurement error3 % o.r.
4 % o.r.
5 % o.f.s.
(depending on chosen option of ordering feature "Calibration flow")

Measuring range0.5 to 910 kg/h (1.1 to 2002 lb/h)
0.5 to 1365 kg/h (1.1 to 3003 lb/h)
(for air, depending on chosen option of ordering feature "Calibration flow")
Max. process pressurePN 40, Class 300
Medium temperature range–40 to +100 °C (–40 to +212 °F)
Ambient temperature range–40 to +60 °C (–40 to +140 °F)
Transmitter housing materialAlSi10Mg, coated
Degree of protectionIP66/67, type 4X enclosure
Display/Operation4‐line display with push Buttons
Configuration via local display and operating tools possible
Outputs4‐20 mA HART (active)
Pulse/frequency/switch output (passive)
InputsStatus input
Digital communicationHART
Power supplyDC 18 to 30 V
Hazardous area approvalsATEX, IECEx, cCSAus
Metrological approvals and certificatesCalibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR
Pressure approvals and certificatesPED, CRN