Max. measurement errorVolume flow (standard): ±0.5 % o.r.± 1 mm/s (0.04 in/s)
Measuring range4 dm³/min to 9600 m³/h (1 gal/min to 44 000 gal/min)
Medium temperature rangeLiner material PFA: –20 to +150 °C (–4 to +302 °F)
Liner material PTFE: –40 to +130 °C (–40 to +266 °F)
Max. process pressurePN 40, Class 300, 20K
Wetted materialsLiner: PFA; PTFE
Electrodes: 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); Tantalum; Platinum; Titanium
Measuring principleElectromagnetic
Product headlineFlowmeter for basic process applications with easy-to-use operation concept.
Dedicated to chemical and process applications with corrosive liquids.
Sensor featuresDiverse applications – wide variety of wetted materials. Energy-saving flow measurement – no pressure loss due to cross section constriction. Maintenance-free – no moving parts.
Nominal diameter: max. DN 600 (24"). All common Ex approvals. Liner made of PTFE or PFA.
Transmitter featuresOptimum usability – operation with mobile devices and SmartBlue app or display with touch screen. Simple, time-saving commissioning – guided parameterization in advance and in the field. Integrated verification – Heartbeat Technology.
System integration with HART, Modbus RS485. Flexible operation with app and optional display.
Nominal diameter rangeDN 15 to 600 (½ to 24")
Wetted materialsLiner: PFA; PTFE
Electrodes: 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); Tantalum; Platinum; Titanium
Measured variablesVolume flow, conductivity, mass flow
Max. measurement errorVolume flow (standard): ±0.5 % o.r.± 1 mm/s (0.04 in/s)
Measuring range4 dm³/min to 9600 m³/h (1 gal/min to 44 000 gal/min)
Max. process pressurePN 40, Class 300, 20K
Medium temperature rangeLiner material PFA: –20 to +150 °C (–4 to +302 °F)
Liner material PTFE: –40 to +130 °C (–40 to +266 °F)
Ambient temperature rangeFlange material carbon steel: –10 to +60 °C (+14 to +140 °F)
Flange material stainless steel: –40 to +60 °C (–40 to +140 °F)
Sensor housing materialDN 15 to 300 (½ to 12""): AlSi10Mg, coated
DN 350 to 600 (14 to 24""): Carbon steel with protective varnish
Transmitter housing materialAlSi10Mg, coated
Degree of protectionStandard: IP66/67, Type 4X enclosure
Display/Operation2.4" LCD display with touch & auto rotate; Configuration and operation via SmartBlue App (Bluetooth) possible
Outputs4-20 mA HART (active/passive), Pulse/frequency/switch output
Modbus RS485, 4-20 mA
Digital communicationHART, MODBUS RS485
Power supplyDC 24 V
AC 100 to 230 V
AC 100 to 230 V / DC 24 V (non-hazardous area)
Hazardous area approvalsATEX, IECEx, cCSAus, EAC, NEPSI, INMETRO, KC, JPN, UK Ex
Metrological approvals and certificatesCalibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation)
Pressure approvals and certificatesPED, CRN
Material certificates3.1 material
Hygienic approvals and certificatesDriking water approvals: ACS; NSF61, WRAS