Endress+Hauser 5P1B Proline Promag P 100 electromagnetic flowmeter

Write a Review

REQUEST A QUOTE

Endress+Hauser 5P1B Proline Promag P 100 electromagnetic flowmeter

Max. measurement errorVolume flow (standard): ±0.5 % o.r.± 1 mm/s (0.04 in/s)
Volume flow (option): ±0.2 % o.r. ± 2 mm/s (0.08 in/s
Measuring range4 dm³/min to 9600 m³/h (1 to 44 000 gal/min)
Medium temperature rangeLiner material PFA: –20 to +150 °C (–4 to +302 °F)
Liner material PFA high‐temperature: –20 to +180 °C (–4 to +356 °F)
Liner material PTFE: –40 to +130 °C (–40 to +266 °F)
Max. process pressurePN 40, Class 300, 20K
Wetted materialsLiner: PFA, PTFE
Electrodes: 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); Tantalum; Platinum; Titanium
Duplex 1.4462 (UNS S31803)
Measuring principleElectromagnetic
Product headlineThe flowmeter for highest medium temperatures with an ultra-compact transmitter.
Dedicated to chemical and process applications with corrosive liquids and high medium temperatures.
Sensor featuresDiverse applications – wide variety of wetted materials. Energy‐saving flow measurement – no pressure loss due to cross section constriction. Maintenance‐free – no moving parts.
Nominal diameter: max. DN 600 (24"). All common Ex approvals. Liner made of PTFE or PFA.
Transmitter featuresSpace‐saving transmitter – full functionality on the smallest footprint. Time‐saving local operation without additional software and hardware – integrated web server. Integrated verification – Heartbeat Technology.
Robust, ultra-compact transmitter housing. Local display available.
Nominal diameter rangeDN 15 to 600 (½ to 24")
Wetted materialsLiner: PFA, PTFE
Electrodes: 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); Tantalum; Platinum; Titanium
Duplex 1.4462 (UNS S31803)

Measured variablesVolume flow, conductivity, mass flow, corrected volume flow, corrected conductivity
Max. measurement errorVolume flow (standard): ±0.5 % o.r.± 1 mm/s (0.04 in/s)
Volume flow (option): ±0.2 % o.r. ± 2 mm/s (0.08 in/s
Measuring range4 dm³/min to 9600 m³/h (1 to 44 000 gal/min)
Max. process pressurePN 40, Class 300, 20K
Medium temperature rangeLiner material PFA: –20 to +150 °C (–4 to +302 °F)
Liner material PFA high‐temperature: –20 to +180 °C (–4 to +356 °F)
Liner material PTFE: –40 to +130 °C (–40 to +266 °F)
Ambient temperature rangeFlange material carbon steel: –10 to +60 °C (+14 to +140 °F)
Flange material stainless steel: –40 to +60 °C (–40 to +140 °F)
Sensor housing materialDN 15 to 300 (½ to 12"): AlSi10Mg, coated
DN 350 to 600 (14 to 24"): Carbon steel with protective varnish
Transmitter housing materialAlSi10Mg, coated
Degree of protectionIP66/67, type 4X enclosure
Display/Operation4‐line backlit display available (no local operation)
Configuration via web browser and operating tools possible
Outputs4‐20 mA HART (active)
Pulse/frequency/switch output (passive)
InputsNone
Digital communicationHART, PROFIBUS DP, Modbus RS485, EtherNet/IP, PROFINET
Power supplyDC 20 to 30 V
Hazardous area approvalsATEX, IECEx, cCSAus, INMETRO
Other approvals and certificates
Product safetyCE, C-Tick
Metrological approvals and certificatesCalibration performed on accredited calibration facilities (acc. to ISO/IEC
17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation)
Marine approvals and certificatesLR approval, DNV approval, ABS approval, BV approval
Pressure approvals and certificatesPED, CRN
Material certificates3.1 material
Hygienic approvals and certificatesDrinking water approval: ACS, NSF 61, WRAS
Returns Policy

Related Products

Top