Endress+Hauser
Endress+Hauser 5H2B Proline Promag H 200 electromagnetic flowmeter
REQUEST A QUOTE
Max. measurement error | Volume flow: ±0.5 % o.r. ± 2 mm/s (0.08 in/s) |
---|---|
Measuring range | 0.06 dm³/min to 300 m³/h (0.015 to 80 gal/min) |
Medium temperature range | –20 to +150 °C (–4 to +302 °F) |
Max. process pressure | PN 40, Class 150, 20K |
Wetted materials | Liner: PFA Electrodes: 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); Tantalum; Platinum Process Connections: stainless steel, 1.4404 (F316L); PVDF; PVC adhesive sleeve Seals: O-ring seal (EPDM, FKM, Kalrez), aseptic molded seal (EPDM, FKM, silicone) Grounding Rings: stainless steel, 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); tantalum |
Measuring principle | Electromagnetic |
Product headline | Flowmeter for smallest flow rates with genuine loop-powered technology. Dedicated to the measurement of the smallest flow quantities. |
Sensor features | Flexible installation concept – numerous process connections. Energy‐saving flow measurement – no pressure loss due to cross section constriction. Maintenance‐free – no moving parts. Liner made of PFA. Sensor housing made of stainless steel. Various electrode materials available. |
Transmitter features | Convenient device wiring – separate connection compartment. Safe operation – no need to open the device due to display with touch control, background lighting. Integrated verification – Heartbeat Technology. Loop-powered technology. Robust dual-compartment housing. Plant safety: worldwide approvals (SIL, Haz. area). |
Nominal diameter range | DN 2 to 25 (¹⁄₁₂ to 1") |
Wetted materials | Liner: PFA Electrodes: 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); Tantalum; Platinum Process Connections: stainless steel, 1.4404 (F316L); PVDF; PVC adhesive sleeve Seals: O-ring seal (EPDM, FKM, Kalrez), aseptic molded seal (EPDM, FKM, silicone) Grounding Rings: stainless steel, 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); tantalum |
Measured variables | Volume flow, mass flow |
Max. measurement error | Volume flow: ±0.5 % o.r. ± 2 mm/s (0.08 in/s) |
Measuring range | 0.06 dm³/min to 300 m³/h (0.015 to 80 gal/min) |
Max. process pressure | PN 40, Class 150, 20K |
Medium temperature range | –20 to +150 °C (–4 to +302 °F) |
Ambient temperature range | –40 to +60 °C (–40 to +140 °F) |
Sensor housing material | 1.4301 (304), corrosion resistant |
Transmitter housing material | AlSi10Mg, coated |
Degree of protection | IP66/67, type 4X enclosure |
Display/Operation | 4‐line backlit display with touch control (operation from outside) Configuration via local display and operating tools possible Remote display available |
Outputs | 4‐20 mA HART (passive) Pulse/frequency/switch output (passive) |
Inputs | None |
Digital communication | HART, PROFIBUS PA, FOUNDATION Fieldbus |
Power supply | DC 18 to 35 V (4‐20 mA HART with/without pulse/frequency/switch output) |
Hazardous area approvals | ATEX, IECEx, cCSAus, INMETRO, NEPSI, EAC, JPN, UK Ex |
Product safety | CE, C-Tick |
Functional safety | Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511 |
Metrological approvals and certificates | Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025) Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation) |
Pressure approvals and certificates | CRN |
Material certificates | 3.1 material |
Returns Policy